Model Number 48-40-00
Device Problem Failure to Sense (1559)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/13/2019
Event Type  malfunction  
Manufacturer Narrative
Patient information: there was no patient involvement.Livanova (b)(4) manufactures the s5 system.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate and tested the device intensively.The service representative could not reproduce the issue.He found that the system was working within specification.Subsequent functional verification testing was completed without further issues and the unit was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that the level sensor alarm did not triggered the reaction of the electrical clamp during procedure.The level dropped down and the perfusionist clamped the line, thus reaching the proper level again.The procedure could be completed without further issue.There was no report of patient injury.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: the affected device has been requested back to the manufacturer site for a detailed investigation.During tests an error code was detected.This error does not allow the clamp to properly close at the level alarm.Blood spots and dried saline solution have been detected into the clamp.Fluid ingress issue has been identified to be the root cause of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8780520
MDR Text Key150700515
Report Number9611109-2019-00532
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number48-40-00
Was the Report Sent to FDA? No
Date Manufacturer Received10/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-