Failure analysis was performed on the main board.Visual inspection: no anomalies - main board found to be clean and in good condition.Benchtop analysis: assembled into a golden unit.Ran on automated test console.Failed atrial blanking and atrial sensitivity at all tested levels (1mv, 2mv, 5mv, and 10mv).Benchtop test using sigma pace 1000 analyzer confirmed atrial sensitivity failure at all levels tested.Ventricular sensitivity found to be within spec.Engineering evaluation and troubleshooting found problem related to a capacitor.Replaced capacitor with a known good capacitor fix the problem.Benchtop test using sigma pace 1000 analyzer confirmed atrial sensitivity pass all tests.Re-run the unit through automated test with full test ¿ all passed.Out of circuit measurement of original capacitor removed from the board found the capacitance within 10% tolerance spec.This confirmed no defect of the original capacitor.Replaced the original capacitor back on the main board.Re-run the unit through benchtop test and automated test ¿ all passed.Based on this analysis data, the root cause is more likely related to capacitor terminal contamination on the board.Confirmed customer complaint.Device failed atrial blanking and atrial sensitivity.Failure caused by capacitor terminal contamination.If information is provided in the future, a supplemental report will be issued.
|