Siemens has completed an investigation of the reported event.The root cause was determined to be a user error.Siemens has concluded that the cause of this event was the introduction of ferromagnetic pieces into the mr examination room and therefore a user error.Due to the strong magnetic field, particular safety measures have to be adhered to in order to prevent injuries.Therefore, the corresponding magnetom operator manual provides clear instructions and warnings regarding both magnetic field hazards and training of personnel with regards to mr safety.However, the responsibility to instruct personnel and patients who have access to the mr examination room about magnetic field hazards lies with the customer.The manuals state that only equipment specified or recommended for use in the controlled area (mr examination room) shall be used.The introduction of ferromagnetic objects into the magnetic field is contrary to the statements given in the operating instructions.Furthermore, special warning signs are posted at the entrance of the controlled access area (magnet room).The manufacturer is not considering further actions resulting from this event as no systematic error has been recognized.
|