SIEMENS HEALTHCARE GMBH MAGNETOM VERIO DOT; SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING Back to Search Results
Model Number 10684333
Device Problem Use of Device Problem (1670)
Patient Problems Bone Fracture(s) (1870); Head Injury (1879)
Event Date 07/12/2019
Event Type  Injury  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that an adverse event occurred while operating the magnetom verio dot system.A magnetic trolley was brought into the examination room and attracted to the magnet resulting in injury to the patient.The patient was immediately moved to a ct scanning room and transferred to the emergency department.It was reported that the patient suffered a rib fracture and injury to the head.Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a user error.Siemens has concluded that the cause of this event was the introduction of ferromagnetic pieces into the mr examination room and therefore a user error.Due to the strong magnetic field, particular safety measures have to be adhered to in order to prevent injuries.Therefore, the corresponding magnetom operator manual provides clear instructions and warnings regarding both magnetic field hazards and training of personnel with regards to mr safety.However, the responsibility to instruct personnel and patients who have access to the mr examination room about magnetic field hazards lies with the customer.The manuals state that only equipment specified or recommended for use in the controlled area (mr examination room) shall be used.The introduction of ferromagnetic objects into the magnetic field is contrary to the statements given in the operating instructions.Furthermore, special warning signs are posted at the entrance of the controlled access area (magnet room).The manufacturer is not considering further actions resulting from this event as no systematic error has been recognized.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNETOM VERIO DOT
Type of Device
SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
henkestrasse 127
erlangen, 91052
GM  91052
MDR Report Key8792728
MDR Text Key151106505
Report Number3002808157-2019-89627
Device Sequence Number1
Product Code LNH
Combination Product (y/n)N
PMA/PMN Number
K121434
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 07/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number10684333
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age50 YR
Patient Weight60
-
-