Catalog Number 08K28-12
Device Problem High Test Results (2457)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Correction/removal number: 3002809144-06/4/19-006-r.A product recall letter was issued to all architect bnp customers who have received calibrator or control lots still within dating.The letter informs the customer of the issue regarding a time dependent stability issue of the architect bnp calibrators and controls that may lead to controls out of range and shift in control and patient results.The letter informs the customer all architect bnp calibrators and controls will have shortened expiration dates of 165 days from the date of manufacture.The letter instructs the customer to discontinue use of the lots which are beyond the 165 day expiration and destroy any remaining inventory.The cause of the shift is instability of the architect bnp calibrators and controls.
 
Event Description
The customer observed abbott controls shift high and out of range while using architect bnp control lot 44k82818.There was no report of incorrect patient results or impact to patient management.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT BNP CONTROLS
Type of Device
ARCHITECT BNP CONTROLS
Manufacturer (Section D)
ABBOTT GERMANY
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GERMANY
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key8796741
MDR Text Key151345456
Report Number3002809144-2019-00439
Device Sequence Number1
Product Code JJX
UDI-Device Identifier00380740145262
UDI-Public00380740145262
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 07/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/15/2019
Device Catalogue Number08K28-12
Device Lot Number44K82818
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction Number3002809144-06/4/19-006R
Patient Sequence Number1
Treatment
ARCHITECT I1000 SR, LIST 01L86-97; SERIAL (B)(4)
-
-