Model Number CI-1600-04 |
Device Problems
Output Problem (3005); Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/01/2019 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing sound quality issues.Programming adjustments were made, however, the issue did not resolve.Revision surgery is scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed top cover silicone slices and the electrode was severed near the fantail and array prior to receipt.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented some electrical tests from being performed.The device passed the electrical and mechanical tests performed.The scanning electron microscopy analysis of the array noted no anomalies found.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|