The log file evaluation revealed the following: after approximately 10 minutes into the procedure the supervisor function of the software detected failure of flow sensor calibration as well as a pressure sensor failure.The device shut down automatic ventilation to protect the patient from potentially harmful output.The corresponding alarms expiratory/inspiratory flow sensor not calibrated, pressure sensor failure and ventilator failure were posted.Patient support was continued in manual ventilation as reported.The recorded error codes indicate that the root cause for the deviation was an incorrect reference voltage for ad conversion on a certain pcb that processes the sensor signals.The particular pcb was replaced, the device was tested and is back in use.Dräger finally concludes that the device responded as designed to a deviation of that nature.When the flow and pressures sensor signals are rated as invalid during an ongoing ventilation episode this is a potential risk of generating harmful output.Thus, the safety concept defines a shut down of automatic ventilation in such a situation.As confirmed for the particular procedure, manual ventilation will further be possible.The overall ffr for the particular pcb is within the expected range of the respective risk assessment and thus accepted.
|