• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number EL5ML
Device Problem Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/05/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # t92z6e.Investigation summary: the analysis results found that the el5ml device was returned with no damage in the external components.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed 9 conforming clips.Upon testing, the jaws opened and closed without any difficulties.In addition, the device locked out as intended.No conclusion could be reached as to what may have caused the reported incident.The reported complaint could not be confirmed.A manufacturing record evaluation was performed for the finished device lot/batch t92z6e number, and no non-conformances were identified.Attempts are being made to obtain the following information: can you please confirm how the device was removed from the tissue/vessel it was on when it would not open? was the device able to be opened? if yes, how were the jaws opened? or was the device cut off of tissue/vessel? to date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported that during a laparoscopic cholecystectomy, the doctor deployed the clip applier but the handle would not return to the open position.A second like device was used to complete the procedure.There were no patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key8804222
MDR Text Key151740630
Report Number3005075853-2019-20473
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2024
Device Catalogue NumberEL5ML
Device Lot NumberT92Z6E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/20/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/11/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-