• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GENTHERM MEDICAL, LLC HEMOTHERM; HEATER / COOLER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GENTHERM MEDICAL, LLC HEMOTHERM; HEATER / COOLER Back to Search Results
Model Number 400CE
Device Problem Power Problem (3010)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/26/2019
Event Type  malfunction  
Manufacturer Narrative
Gentherm received a report (b)(6) that the hemotherm stopped working during a procedure.No patient harm or injury occurred.No other information has been provided.
 
Event Description
Hemotherm 400ce stopped working during a procedure.The unit alerted with an "ee03" alarm.It was reported that the patient did not have an adverse event.
 
Manufacturer Narrative
Device evaluated and determined to have a defective thermistor.The thermistor was replaced to correct.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOTHERM
Type of Device
HEATER / COOLER
Manufacturer (Section D)
GENTHERM MEDICAL, LLC
12011 mosteller
cincinnati OH 45241
Manufacturer (Section G)
GENTHERM MEDICAL, LLC
12011 mosteller
cincinnati OH 45241
Manufacturer Contact
christina miracle
12011 mosteller
cincinnati, OH 45241
8009897373
MDR Report Key8809660
MDR Text Key152085281
Report Number1516825-2019-00009
Device Sequence Number1
Product Code DWC
UDI-Device Identifier10613031860221
UDI-Public(01)10613031860221
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122813
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number400CE
Device Catalogue Number86022
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/19/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-