• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRULIGHT TRIO TUSM; SURGICAL LIGHT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRULIGHT TRIO TUSM; SURGICAL LIGHT Back to Search Results
Model Number 1574851
Device Problems Mechanical Problem (1384); Detachment of Device or Device Component (2907)
Patient Problem No Patient Involvement (2645)
Event Date 06/25/2019
Event Type  malfunction  
Manufacturer Narrative
A trumpf medical service technician inspected the surgical light system and found that the snap ring was loose.The snap ring was replaced and the light functioned as designed.A field corrective action has been initiated by trumpf medical and reported to the fda as a result of investigations into similar events (recall number z-563-2016).The investigations have found that improper installation or servicing of the circlip in this joint can result in the slipping down or falling of the spring arm and light head components.Based on this information, no further action is required.
 
Event Description
A customer reported that the spring arm popped out and dropped down on a trumpf medical surgical light system.An inspection of the device revealed that the snap ring came loose, allowing the spring arm to drop an inch and a half.No injuries were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIGHT TRIO TUSM
Type of Device
SURGICAL LIGHT
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
carl-zeiss strasse 7-9
saalfeld, 07318
GM  07318
Manufacturer Contact
steffen ulbrich
carl-zeiss strasse 7-9
saalfeld, 07318
GM   07318
MDR Report Key8823117
MDR Text Key152088158
Report Number9681407-2019-00018
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102758
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Biomedical Engineer
Remedial Action Repair
Type of Report Initial
Report Date 06/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1574851
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/25/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/19/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-