MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 48-40-00
Device Problem No Audible Alarm (1019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/01/2019
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova (b)(4) manufactures the s5 system.The incident occurred in (b)(6) a livanova field service representative was dispatched to the facility to investigate and was able to confirm the reported issue.Investigation of the issue is ongoing.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that on a s5 system the alarm is not audible when turned down to 20%, whereas the alarm is fine on 100% there was no report of patient injury.
 
Manufacturer Narrative
H.10: a video was provided to livanova deutschland.Results of the investigation revealed that, only after powering off and on the system, the volume did not hold the same setting.This is a software feature.The user should decide about the default value in the system settings of the system menu.This value would have been stored and after every start-up of the machine or panel, it would have been resumed.The device was working within specification and no problem could be identified.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key8836349
MDR Text Key159518513
Report Number9611109-2019-00578
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 10/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number48-40-00
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-