MAUDE Adverse Event Report: COOK INC POLYETHYLENE ANGIOGRAPHIC CATHETER; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC POLYETHYLENE ANGIOGRAPHIC CATHETER; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number G00174
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Patient Involvement (2645)
Event Date 07/22/2019
Event Type  malfunction  
Manufacturer Narrative
Occupation: unknown healthcare professional.Pma/510(k) number: pre-amendment.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event Description
As reported, a hair was found inside the package of a polyethylene angiographic catheter.The device did not make patient contact.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation - evaluation reviews of the complaint history, device history record, documentation, manufacturer¿s instructions, and quality control procedures, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device revealed a hair-like fiber present inside the packaging.Additionally, a document-based investigation evaluation was performed.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Furthermore, reviews of the manufacturer¿s instructions and quality control procedures were conducted, and no gaps were discovered.Based on the information provided and the examination of the returned product, investigation has concluded that the failure mode can be attributed to the manufacturing and packaging of the device.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYETHYLENE ANGIOGRAPHIC CATHETER
Type of Device
DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key8842695
MDR Text Key152950421
Report Number1820334-2019-01872
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00827002001749
UDI-Public(01)00827002001749(17)210605(10)8852653
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/05/2021
Device Model NumberG00174
Device Catalogue NumberP6.7-38-100-P-4S-GPC
Device Lot Number8852653
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2019
Date Manufacturer Received09/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-