|
Catalog Number 1049 |
Device Problem
Disconnection (1171)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/08/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
The complaint is reported as: "mask oxygen connection does not connect properly to oxygen tubing.Tubing easily falls off mask.New mask in use, oxygen tubing remains consistent.".
|
|
Manufacturer Narrative
|
Qn#(b)(4).The sample was returned for evaluation.It was also found that additional oxygen tubing that is not part of the finished good was re turned.During the visual exam it was observed that the oxygen tubing is easily disconnected from the grommet.The grommet that is assembled on the received mask was found within dimensional specification.A device history record review was performed and no relevant findings were identified.Based on the investigation performed, the reported complaint could not be confirmed.
|
|
Event Description
|
The complaint is reported as: "mask oxygen connection does not connect properly to oxygen tubing.Tubing easily falls off mask.New mask in use, oxygen tubing remains consistent.".
|
|
Search Alerts/Recalls
|
|
|