Patient's weight was unavailable.Device lot number and expiration date are not available from the facility.The device was discarded by the user.Device manufacture date is dependent on the device lot number, thus is unavailable.The device was discarded.Therefore, no device evaluation can be performed.
|
A philips representative reported that a cardiac lead management procedure commenced to remove a non-functional/fractured right ventricular (ev) implantable cardioverter defibrillator (icd) lead.A spectranetics lead locking device (lld) was utilized as the traction platform for the extraction.The lld snapped at the rv lead fracture area, which was at the proximal end of the coil.The patient's blood pressure dropped at that time.Rescue efforts, including cpr and sternotomy, revealed that the superior vena cava (svc) was torn during the extraction (please see mdr 1721279-2019-00136, as the svc tear is not attributed to the lld device).Rescue efforts commenced including sternotomy, and the lead was removed during open heart.Note: 4076 right atrial (ra) lead, rep believes, was taken out during the open heart along with the 0292 right ventricular (rv) lead, due to being potentially dislodged.Original intention was to just take out 0292 rv lead.The patient survived the procedure.
|