• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUROPACE, INC. NEUROPACE RNS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number DL-344-10-K
Device Problems Signal Artifact/Noise (1036); Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/22/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The explanted lead was not returned to neuropace for analysis.However review of the rns neurostimulator ecog and lead impedance data confirmed that the ecog data were suggestive of a potential lead break.
 
Event Description
Review of the rns neurostimulator ecog data indicated a lead artifact on the right depth lead.Therapy and detection were disabled on the affected lead.No information regarding seizure-related trauma or other potential causal factors resulting in the lead break were provided by the treating center.On (b)(6) 2018 the affected lead was explanted and a new lead was implanted.No complications were reported during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key8853541
MDR Text Key211498049
Report Number3004426659-2019-00040
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005014
UDI-Public010085554700501417190620
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberDL-344-10-K
Device Catalogue Number1007607
Device Lot Number22331-1-1-1
Was Device Available for Evaluation? No
Date Manufacturer Received01/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age44 YR
-
-