Model Number 1458Q/92 |
Device Problems
Failure to Capture (1081); Loose or Intermittent Connection (1371); Pocket Stimulation (1463); Failure to Disconnect (2541); Device Dislodged or Dislocated (2923)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/22/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that when patient presented to clinic as a new patient, device interrogation revealed loss of capture on lv lead along with pocket stimulation from pacing.X-ray revealed lv lead dislodgement.Lv pacing was programmed off.Lead revision was planned.The patient was stable.
|
|
Event Description
|
New information received notes that the when patient presented for lead revision, the lead was not able to be re-inserted as guidewire could not be removed from the distal end of lead.It was further noted that the lead was not attached to a suture sleeve in the pocket and that the suture sleeve was still tied down by itself.The old left ventricular(lv) lead was explanted and replaced.The patient was stable.
|
|
Manufacturer Narrative
|
The damage found was sustained during the surgical procedure.The lead was otherwise normal.
|
|
Search Alerts/Recalls
|