Siemens healthcare diagnostics, inc.Has confirmed that the atellica ch 930 module may report incorrect serum indices for hemolysis (h) and lipemia (l) when using the un_c test as a donor method, and the hemolysis icterus lipemia (hil) function is enabled.Due to bubbles in the reaction cuvette, there is a potential for falsely depressed 'h' results (e.G.H index of 3 or higher, reported as less than or equal to 1 ) at a rate of 0.3-1% and falsely elevated 'l' results (e.G.L index of 0 reported as more than or equal to 1) at a rate of 2.9%.In all cases of falsely depressed h, l is elevated more than or equal to1.For samples not using un_c as the donor method the indices are not impacted.An urgent medical device correction (umdc) asi19-02.A.Us was sent to us customers and an urgent field safety notice (ufsn) asi19-02.A.Ous was sent to ous customers in july 2019.The umdc and ufsn advise customers who have multiple atellica ch 930 analyzers to generate un_c results on a different analyzer than the hil tests.For customers who have only one atellica ch 930 analyzer when un_c is part of the test order and alt and ast are not, hil tests should be ordered separately, or alternatively hil ordering can be turned off.
|
One discordant, falsely elevated lipemia (l) index was obtained on a single patient sample when using atellica ch 930 module.The initial index was flagged incorrectly and was not reported to the physician(s).Observation found the sample to be visibly not lipemic.The repeat index from another atellica ch 930 analyzer (serial number (b)(4)) was correct and reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated lipemia result.
|