• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC KINCISE CUP-ADAPTER-PINNACLE STRAIGHT; IMPACTOR, ADAPTERS, BATTERY, BATTERY CHARGER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC KINCISE CUP-ADAPTER-PINNACLE STRAIGHT; IMPACTOR, ADAPTERS, BATTERY, BATTERY CHARGER Back to Search Results
Catalog Number 1011-01-101
Device Problem Crack (1135)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/11/2019
Event Type  malfunction  
Manufacturer Narrative
The reporter's complete address was not provided.The actual device was returned for evaluation.Quality engineering evaluated the device and it was determined that the reported condition that the adapter device was cracked on the crossbar was confirmed.An assessment was performed on the device which found that the t-handle was broken.It was determined that the device was most likely side loaded when used with the impactor which is improper device use.The assignable root cause was determined to be due to user error.A review of the device history was performed and no non-conformances were detected related to the reported condition.If additional information should become available, a supplemental medwatch report will be submitted accordingly.(b)(4).
 
Event Description
It was reported that during an unspecified surgical procedure it was observed that the adapter device broke in half near the "t" crossbar.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization.During in-house engineering evaluation it was observed that the t-handle on the device was cracked on the crossbar and broken.This event did not occur during surgery.There was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KINCISE CUP-ADAPTER-PINNACLE STRAIGHT
Type of Device
IMPACTOR, ADAPTERS, BATTERY, BATTERY CHARGER
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
TECHTRONIC INDUSTRIES MEDICAL
1428 pearman dairy rd
anderson SC 29625
Manufacturer Contact
kara ditty-bovard
4500 riverside drive
palm beach gardens, FL 33410
6103142063
MDR Report Key8890055
MDR Text Key154556252
Report Number1045834-2019-54287
Device Sequence Number1
Product Code GEY
UDI-Device Identifier00850915006099
UDI-Public850915006099
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 06/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1011-01-101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2019
Date Manufacturer Received08/08/2019
Date Device Manufactured06/18/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-