Model Number DL-344-3.5-K |
Device Problems
Signal Artifact/Noise (1036); Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/17/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).The explanted lead was not returned to neuropace for analysis, however review of the lead impedance and ecog data was performed and the data is consistent with a lead break.Unable to determine root cause.
|
|
Event Description
|
Neuropace became aware of a possible lead break based on review of the patient's ecog data in (b)(6) 2019 and the treating center was notified.The rns system remained implanted with stimulation and detection enabled.The center performed confirmatory testing in clinic; however the testing was limited due to the clear signal change and abnormal impedance measurements.The right depth lead was subsequently explanted and a new lead implanted on (b)(6) 2019 during a routine neurostimulator replacement.
|
|
Search Alerts/Recalls
|