• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUROPACE, INC. NEUROPACE RNS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number DL-344-3.5-K
Device Problems Signal Artifact/Noise (1036); Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/17/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).The explanted lead was not returned to neuropace for analysis, however review of the lead impedance and ecog data was performed and the data is consistent with a lead break.Unable to determine root cause.
 
Event Description
Neuropace became aware of a possible lead break based on review of the patient's ecog data in (b)(6) 2019 and the treating center was notified.The rns system remained implanted with stimulation and detection enabled.The center performed confirmatory testing in clinic; however the testing was limited due to the clear signal change and abnormal impedance measurements.The right depth lead was subsequently explanted and a new lead implanted on (b)(6) 2019 during a routine neurostimulator replacement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key8898197
MDR Text Key154540307
Report Number3004426659-2019-00037
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005021
UDI-Public010085554700502117150911
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 07/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberDL-344-3.5-K
Device Catalogue Number1007606
Device Lot Number19102-1-1-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/17/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age38 YR
-
-