• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESYNT INCORPORATED OMNICELL I.V.STATION; PHARMACY COMPOUNDING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESYNT INCORPORATED OMNICELL I.V.STATION; PHARMACY COMPOUNDING DEVICE Back to Search Results
Model Number I.V. STATION
Device Problems Incorrect Measurement (1383); Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/18/2019
Event Type  malfunction  
Event Description
On july 18, 2019 an i.V.Station device passed multiple underdosed and inadequately reconstituted vials used for final drug preparation.Sandoz vials used for cefazolin preparations were not effectively weighed during preparation, causing vials to be underfilled with diluent, not fully reconstituted after shaking and outside of the allowed percentage of error for vial preparation.All vials were identified and there are no known adverse patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNICELL I.V.STATION
Type of Device
PHARMACY COMPOUNDING DEVICE
Manufacturer (Section D)
AESYNT INCORPORATED
51 pennwood place
warrendale PA 15086
Manufacturer (Section G)
AESYNT INCORPORATED
51 pennwood place
warrendale PA 15086
Manufacturer Contact
david vanella
500 cranberry woods drive
suite 400
cranberry township, PA 16066
7247418115
MDR Report Key8901874
MDR Text Key159198528
Report Number3011278888-2019-00088
Device Sequence Number1
Product Code NEP
UDI-Device Identifier08056477790007
UDI-Public(01)08056477790007(11)190118
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 08/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberI.V. STATION
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/08/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-