Age at time of event: 18 years or older device evaluated by mfr: the tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The shaft showed 2 kinks located at the occ handle and 177.5cm from the tip.There was coating peeling from the 177.5 location.The occ handle was connected to the ffr link for signal verification.The signal was present as designed.The occ handle cap was loosened to remove the wire.There was no issue with removing the wire.The sensor port showed no residue of body fluids.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The complaint of a kink was confirmed.The pressure wire was connected to the analysis support test bench and all applicable data was correct as designed, there was no difficulty in connecting the wire to the test equipment.The coefficient was confirmed to be in specification.
|