MAUDE Adverse Event Report: RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Failure to Auto Stop (2938)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
The device involved in the reported incident is not available for evaluation.An investigation has been initiated based on the reported information.
 
Event Description
The customer also reports that in the past, too, other perforators (which are not broken) have come too deep, failed to disengage.
 
Manufacturer Narrative
Perforator was received for evaluation: dhr - all assembly process steps, cleaning, functional testing, labeling, packaging, and sterilization process steps were followed and completed with zero non-conformances.- lot j1839j ,device identifier # 10886704005100, manufacturing date of 07-dec-2018 and expiration date of 30-nov-2023 failure analysis - the returned unit was found to meet all acceptance criteria.The complaint could not be verified through failure analysis.The cause cannot be confirmed, as no issues relevant to the failure were noted from the dhr review, trending, or failure analysis, and proper finished goods testing was performed prior to release.Between 05nov2019 and 30jun2020, approximately 2,200 mdrs submitted electronically by integra lifesciences via trackwise, integra's complaint handling system, were not received by cdrh due to a computer system issue.Within this time period, an error with integra's middleware, which facilitates communications between trackwise and the fda system, caused the complaint records to close and indicate we had received an acknowledgement 3 from the fda when we had not.Integra interpreted the acknowledgement as a successful submission; however, subsequent investigation revealed the acknowledgement 3 received was from our middleware and not from the fda (these acknowledgements have been retained as part of the documentation of the mdr).The malfunction was related to the relocation of the trackwise application to a new data center during the transition of integra's corporate headquarters from plainsboro, nj to princeton, nj.Previously, integra had been successfully receiving acknowledgements 1, 2, and 3 from the fda, and our records reflect these acknowledgements, including the date and time stamps.Capa pr 229048 and nc 20-011 have been opened by integra to further investigate the nonconformance and develop a corrective action plan.The middleware error has been corrected, and integra has filed mdrs since the correction and verified that the appropriate acknowledgements have been received from the fda.Integra is resubmitting all impacted mdr reports for the time period 05nov2019 through 30jun2020.Integra lifesciences contacted (b)(6), director of regulatory programs, office of product evaluation and quality and (b)(6), assistant director, mdr team, office of product evaluation and quality on july 8-9, 2020 to report these issues regarding mdr reports.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA 02767
MDR Report Key8906054
MDR Text Key154741659
Report Number1226348-2019-00335
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number261221
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/26/2019
Date Manufacturer Received03/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-