Model Number LNQ11 |
Device Problems
Circuit Failure (1089); Battery Problem (2885)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.Analysis confirmed crystal errors causing telemetry delays.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the implantable cardiac monitor (icm) was explanted due to end of battery life.The device was returned to the manufacturer, analyzed, and tested out of specification.No patient involvement or complications were reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|