• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIG FEM ADPT TORQUE WRENCH; KNEE INSTRUMENT : TORQUE DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIG FEM ADPT TORQUE WRENCH; KNEE INSTRUMENT : TORQUE DEVICES Back to Search Results
Catalog Number 961673
Device Problems Break (1069); Component Missing (2306)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/01/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the torque wrench was missing the washer.The 2.5 tc3 insert trial metal post was bent.The clamp on the fem/tib/sleeve clamp has opened up and does not engage the implants anymore.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: examination of the returned device revealed the black plastic protector component was missing.The investigation findings did not indicate that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIG FEM ADPT TORQUE WRENCH
Type of Device
KNEE INSTRUMENT : TORQUE DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8918048
MDR Text Key155188975
Report Number1818910-2019-101917
Device Sequence Number1
Product Code HXC
UDI-Device Identifier10603295239789
UDI-Public10603295239789
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number961673
Device Lot NumberSO2012909
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2019
Date Manufacturer Received09/12/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-