• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) BD VACUTAINER® NO ADDITIVE (Z) PLUS TUBES; SPECIMEN TRANSPORT AND STORAGE CONTAINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) BD VACUTAINER® NO ADDITIVE (Z) PLUS TUBES; SPECIMEN TRANSPORT AND STORAGE CONTAINER Back to Search Results
Catalog Number 366408
Device Problem Device Markings/Labelling Problem (2911)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/05/2019
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A sample is available for evaluation.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that prior to use it was discovered that the labels were falling off the tubes with a bd vacutainer® no additive (z) plus tubes.The following information was provided by the initial reporter: (6 of 7), it was reported that the labels are coming off the tubes.
 
Manufacturer Narrative
Investigation summary: bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for label lift with the incident lot was observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd has initiated further investigation relating to this issue through a capa #1064141.The investigation is still on-going and improvements will be made as the potential causes of this issue are identified.
 
Event Description
It was reported that prior to use it was discovered that the labels were falling off the tubes with a bd vacutainer® no additive (z) plus tubes.The following information was provided by the initial reporter: (6 of 7).It was reported that the labels are coming off the tubes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® NO ADDITIVE (Z) PLUS TUBES
Type of Device
SPECIMEN TRANSPORT AND STORAGE CONTAINER
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
MDR Report Key8918837
MDR Text Key155225676
Report Number1024879-2019-01492
Device Sequence Number1
Product Code FMH
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup
Report Date 09/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/15/2019
Device Catalogue Number366408
Device Lot Number9074915
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2019
Date Manufacturer Received08/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-