• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) ACCENT DR RF PACEMAKER; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) ACCENT DR RF PACEMAKER; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2210
Device Problems Pacemaker Found in Back-Up Mode (1440); Communication or Transmission Problem (2896)
Patient Problems Discomfort (2330); Palpitations (2467)
Event Date 07/27/2019
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient felt some palpitations and minor discomfort due to occasional unipolar pacing and presented in clinic for routine follow up.Upon interrogation, it was noted that the pacemaker was in backup vvi and the device was nearing eri.The patient was moving to a new home and the cause of the backup vvi status was unknown, but the device image indicated that it was probably caused by an event que overflow (eqo).A firmware download was performed, and the device was restored to its normal function.The patient was in stable condition post interrogation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCENT DR RF PACEMAKER
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8919205
MDR Text Key155188879
Report Number2938836-2019-12679
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734503617
UDI-Public05414734503617
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2013
Device Model NumberPM2210
Device Lot Number3449651
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/07/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/19/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age80 YR
-
-