The initial reporter complained of a questionable elecsys dhea-s result for 1 patient tested on a cobas 8000 e 801 module compared to a beckman coulter, abbott architect, and siemens advia centaur.The dhea-s result from the cobas e801 was >1000 ug/dl with a data flag.The dheas results from the competitor system were: beckman coulter 358.9 ug/dl, abbott architect 366.2 ug/dl, and siemens advia centaur 347.6 ug/dl.The result in question was reported outside of the laboratory to the patient.The cobas e801 serial number is (b)(4).
|
The customer provided two samples for investigation.Investigation of the samples confirmed an immunoglobulin was present that reacts with the reagent and affects the sample results.This type of interference is addressed in the product labeling: "in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design.".
|