Biocompatibility testing to iso 10993 was successfully completed on skin-contacting surfaces of the lifevest device as well as the blue(tm) defibrillation gel.Electrode belt (b)(4) was returned and evaluated at the distributor in accordance with procedures recommended by zoll manufacturing corporation.The evaluation included download data review and incoming functional testing.During the incoming functional testing, the ecg acquisition and pulse delivery circuitry were verified.Upon evaluation, there was gel leaking from the front and rear therapy electrodes.The root cause of the gel leak was excessive force.
|
A us distributor reported that the patient had developed a deep red, itchy, dry, heat-like rash under the back therapy electrodes.It was also reported that the electrode belt was leaking gel from the therapy electrodes.The patient was given a cream from her physician.After using the cream, the irritation improved.
|