NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number DL-344-10-K
Device Problems Signal Artifact/Noise (1036); Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/25/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4) -the explanted lead was not returned to neuropace for analysis.Review of the rns neurostimulator ecogs and lead impedances confirmed that the ecog data were suggestive of a potential lead break.
 
Event Description
In (b)(6) 2017 a review of the ecog data suggested a possible lead break on the right depth lead.In response to the signal changes, adjustments were made to the responsive stimulation settings at that time.The family did not report any information regarding seizure-related trauma or other potential causal factors resulting in the lead break.The family decided to wait to replace the affected lead until the patient underwent a routine device replacement; at which time the affected lead was explanted and a new depth lead was implanted and connected to the rns neurostimulator without complication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key8925817
MDR Text Key155386398
Report Number3004426659-2019-00039
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005014
UDI-Public010085554700501417190510
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberDL-344-10-K
Device Catalogue Number1007607
Device Lot Number22188-1-1-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/25/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient Age29 YR
-
-