Model Number 1458Q/86 |
Device Problem
Difficult to Advance (2920)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/19/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results, method, and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
During an implant procedure, it was noted that the left ventricular (lv) lead had difficulty advancing with the guidewire.The lv lead was replaced to resolve the event.The patient was stable and will continue to be monitored.
|
|
Manufacturer Narrative
|
The reported event of guidewire failed to advance was not confirmed.As received, a complete lead was returned in one piece for analysis.A guidewire could be fully inserted into the lead without any difficulty.Visual inspection and x-ray examination of the lead did not reveal any anomalies.Electrical testing did not reveal any indication of conductor fractures or internal shorts.
|
|