RESPIRONICS CALIFORNIA, INC V200 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number V200
Device Problem Device Emits Odor (1425)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2018.Date of report: 26aug2019.The manufacturer's field service engineer (fse) confirmed the reported electrical burnt smell issue.The customer replaced the defective power supply to address the reported problem.The power supply, 3rd generation, v200 was returned to failure investigator (fi) lab for evaluation.Visual inspection of this power supply did not reveal any signs of damage or contamination.This power supply will be installed into the fi test ventilator and an attempt to boot into the normal ventilation mode will be made while operating on the ac source.The ac power led did activate when power was applied.Cycling the power on and off did not produce any errors.Final results: the test ventilator did boot up and deliver breaths and the test ventilator did not generate any errors during approximately 4 hours of operation.Mdr has been reassessed as reportable after a request from the fda on march 1, 2019 to review complaints from 29 nov 2017 to 29 nov 2018.As this has been reassessed, it will appear to be a late mdr.
 
Event Description
The customer reported 7002 -the unit has a electrical burnt smell.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V200 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key8931168
MDR Text Key155699319
Report Number2031642-2019-07223
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2018
Was the Report Sent to FDA? No
Date Manufacturer Received05/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/17/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse