Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (gelled belt) has been confirmed.Upon investigation the electrode belt failed a therapy electrode recognition test.The cause for the failure was isolated to the dn to rear te cable which was pulled from strain relief, pulling heat shrink off of pulse wires shorting them together.The belt gelled due to the gel fire wires shorting together.The root cause for the strained cable was excessive force.No adverse event resulted from the defective electrode belt.
|