Model Number 801041
Device Problem Increased Pump Speed (1501)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/08/2019
Event Type  malfunction  
Event Description
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the sucker pump ramped up to a high speed with no input from the perfusionist.The pump was restarted and the issue was resolved.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.Per clinical review: the pump was set up without issue with a quarter inch polyvinyl chloride (pvc) tubing, and the occlusion was set accordingly for the procedure on (b)(6) 2019.Once on bypass, the team turned on the pump from the central control monitor (ccm) icon.The roller pump went to max speed, without input on the slider from the user.He automatically turned the pump off, and then tried to turn it back on with the ccm and was able to appropriately increase the speed with the slider.The team was able to use the roller pump without issue the remainder of the case.This occurrence only happened once during the procedure.They have used the roller pump since without issue.There was no harm or blood loss associated with the event.This over-speed event did not delay the continuation of the surgical procedure.
 
Manufacturer Narrative
The field service representative (fsr) performed preventive maintenance (pm) on 30-sep-2019 and observed the unit operated to the manufacturer's specifications.
 
Manufacturer Narrative
Updated block: h6.The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key8944533
MDR Text Key159864003
Report Number1828100-2019-00464
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801041
Device Catalogue Number801041
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-