Model Number 1574850
Device Problems Degraded (1153); Flaked (1246); Material Integrity Problem (2978)
Patient Problem No Patient Involvement (2645)
Event Date 08/06/2019
Event Type  malfunction  
Manufacturer Narrative
A trumpf medical service technician was dispatched to evaluate the light system due to the paint chipping on the light head.The light head was sanded and cleaned.The affected area was sealed to prevent any additional paint chipping, and the device is functioning as designed.The investigation identified steps in the preparation for coating process which can lead to lack of paint adhesion resulting in paint chipping.Additionally, collisions of the light heads and use of unapproved cleaners can further contribute to paint chipping from the light system.The damage to the painted surfaces usually does not develop suddenly but develops over time.The instructions for use state that the devices must be checked for proper condition prior to each use.Damaged devices must not be used.If damage to the surface coating is recognized and removed, there is no danger to the patient.The surface coating process has been updated to include additional testing and verification.Based on this information, no further action is required.
 
Event Description
A customer reported that the paint on the light head was chipping.No injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIGHT DUO TUSM
Type of Device
SURGICAL LIGHT
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
carl-zeiss strasse 7-9
saalfeld, 07318
GM  07318
Manufacturer Contact
steffen ulbrich
carl-zeiss strasse 7-9
saalfeld, 07318
GM   07318
MDR Report Key8950318
MDR Text Key196290647
Report Number9681407-2019-00024
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102758
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Repair
Type of Report Initial
Report Date 08/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1574850
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/06/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-