It was reported that the patient presented in the hospital due to experiencing skipped beats.During device interrogation, noise was noted on the right ventricular (rv) lead.The noise was not reproducible with isometric exercises, but the lead noise was picked up on the egms whenever the patient coughed.There were multiple noise reversion episodes that had occurred on (b)(6) 2019, and on the day of interrogation, (b)(6) 2019.The rv lead noise was causing the device to over-sense on the rv lead, causing inhibited pacing at times for the patient.Programming changes were made.However, the issue was not completely resolved.The patient was scheduled for lead replacement and device exchange procedure, as the pacemaker had less than three months to normal elective replacement indicator (eri).Subsequently, the generator was changed-out electively and the rv lead was capped and replaced on (b)(6) 2019.The patient was stable throughout the event.
|