TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM Back to Search Results
Model Number 6382
Device Problem Device Sensing Problem (2917)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/12/2019
Event Type  malfunction  
Event Description
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the flow sensor was not reading.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.Per clinical review: the team had an issue with the stand alone centrifugal pump during a cpb procedure on (b)(6) 2019.The stand-alone centrifugal pump was primed without issue.About half way through the procedure, the light emitting diode (led) on the display of the pump flow were displaying.035 liters per minute (l/min).The perfusionist changed his flow probe and it was still displaying a.035 number.The end user decided to proceed without exchanging the flow probe, for he was able to see that his revolutions per minute (rpm) were what he expected them to be, and his patients pressures along with venous saturation levels were as expected.He deduced that the display was the concern not the actual flow probe or centrifugal pump, therefore he felt comfortable without exchanging the entire module.There was no delay in the continuation of the surgical procedure.There was no harm or blood loss due to the event.
 
Manufacturer Narrative
The reported complaint was not verifiable.Multiple diligence attempts for part return were unsuccessful so no evaluation could be performed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS CENTRIFUGAL SYSTEM
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key8969792
MDR Text Key159611112
Report Number1828100-2019-00472
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K950739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 12/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/06/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6382
Device Catalogue Number6382
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-