• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE® CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE® CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number 2AF283
Device Problem Material Twisted/Bent (2981)
Patient Problem No Patient Involvement (2645)
Event Date 07/11/2019
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the balloon catheter, 2af283 with lot number 69585, was returned and analyzed.Visual inspection of the balloon catheter showed the luer was broken.Smart chip verification indicated the catheter was used not.Without reprogramming the catheter, it was connected to the console and no system notices were received; catheter was recognized.The catheter passed electrical integrity test and the performance as per specification.In conclusion, the balloon catheter failed the returned product inspection due to a broken luer.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The balloon catheter subsequently tested out of specification per the manufacturer's investigation.
 
Manufacturer Narrative
Correction: evaluation codes.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE® CARDIAC CRYOABLATION CATHETER
Type of Device
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key8974632
MDR Text Key156917882
Report Number3002648230-2019-00646
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
PMA/PMN Number
P100010/S015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 10/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/24/2021
Device Model Number2AF283
Device Catalogue Number2AF283
Device Lot Number69585
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2019
Date Manufacturer Received10/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-