Catalog Number A1059
Device Problems Detachment of Device or Device Component (2907); Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
This mdr is being submitted as part of service and repair remediation activities associated with capa (b)(4).The device was returned to the manufacturer for service and repair.The unit was cleaned per protocol.It was noted upon inspection that the returned unit did not meet all specific functional tests.The set screw in the swivel base was found to be tight.The unit was received with the plunger stud and cap separated from the base.The locking mechanism had both rotational and lateral movement as well as a buildup of residue.New components were added to replace the worn internal parts.The unit was also machined to have heli-coils added to the large starburst threads.A review of the device's manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances related to the reported failure.
 
Event Description
The user facility returned the a1059 mayfield modified skull clamp for service and repair because the plunger stud was removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH 45227
Manufacturer Contact
kimberly shelly
311 enterprise drive
plainsboro, NJ 08536
6099362319
MDR Report Key8975843
MDR Text Key190166754
Report Number3004608878-2019-00891
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/28/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-