• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, DUAL 4 CHANNEL 10CM; SCS LEAD EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, DUAL 4 CHANNEL 10CM; SCS LEAD EXTENSION Back to Search Results
Model Number 3341
Device Problems Disconnection (1171); High impedance (1291)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/27/2019
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer reference number: 1627487-2019-10152, 1627487-2019-10153.It was reported the diagnostics reports during the programming indicated high impedance on the supra orbital lead.X-rays revealed that the lead pulled out of the extension header.As a result, patient underwent surgical intervention on (b)(6) 2019 where in the lead was re-connected to the extension.Issue resolved post-operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, DUAL 4 CHANNEL 10CM
Type of Device
SCS LEAD EXTENSION
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8977297
MDR Text Key156976836
Report Number1627487-2019-10155
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05414734402279
UDI-Public05414734402279
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K930536
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/11/2019
Device Model Number3341
Device Lot Number6152919
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/27/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/11/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3166 SCS LEAD (2)
Patient Outcome(s) Other;
Patient Age32 YR
Patient Weight68
-
-