S7 with axiem was being used during a shunt placement procedure.The axiem was working properly during registration and accuracy confirmation, but box and emitter went to red status about 30 minutes into the case.They reseated the power/communications cable and rebooted the software with no change.Navigation was aborted and the procedure was completed successfully without guidance.Upon inspection of the system, the axiem box and emitter showed green status and worked properly until left on for approx.30 mins when the box and emitter went into red status and would not revert to green.Ts shipped in a replacement axiem box and emitter.Both parts were replaced and the issue has been resolved.
|