Reference: (b)(4).Investigation: x-review dhr.X-inspect returned samples.Analysis and findings: distribution history: this complaint unit was manufactured at csi on 3/29/18 under wo #239015 and shipped on 6/26/18.Manufacturing record review: dhr 239015 was reviewed and no non-conformities, related to the complaint condition, were noted.Incoming inspection review: not applicable.Service history record: no additional service history records found for this unit.Historical complaint review: a review of the attached 2-year complaint history showed similar reported complaint conditions.Most tend to be much older units and damaged.Product receipt: the complaint unit was returned on a repair under warranty.However, based on log 92486, this unit was at csi on 7/25/19.Visual evaluation: visual examination of the complaint unit revealed no physical damage.Functional evaluation: complaint unit was functionally evaluated and found not to function properly.Service & repair confirmed the trigger was stuck.The release lever was not disengaging the locking lever (freeze mode).Root cause : no definitive root cause for this issue could be reliably determined at this time.This unit was fully functional at assembly and in the field for over a year before the issue arose.The repair was an adjustment to the releasing lever to disengage the locking lever from the 'on' position.Correction and/or corrective action the unit was repaired, tested and returned to the customer under warranty.Coopersurgical will continue to monitor this complaint condition for trends.No further corrective action is necessary.No further training required at this time.Was the complaint confirmed? yes.Preventative action activity coopersurgical will continue to monitor this complaint condition for any trends.
|