Model Number 801764 |
Device Problem
Failure to Run on Battery (1466)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/01/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the manufacturer's subsidiary, the battery icon color on the central control monitor (ccm) and the light emitting diode (led) on the front panel of the heart-lung machine (hlm) were both green.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the system would not switch from alternating current (ac) to battery power.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was not verifiable.Multiple diligence attempts for part return were unsuccessful therefore no evaluation could be performed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
Per the manufacturer's subsidiary, the power manager board was replaced, but the issue remained.
|
|