Model Number 1458Q/86
Device Problem Failure to Capture (1081)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Information (3190)
Event Date 08/12/2019
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.Further information was requested, but not received.
 
Event Description
Related manufacturer reference number: 2938836-2019-13436, 2938836-2019-13437.It was reported that the patient presented for an advisory generator change.The implantable cardioverter defibrillator had reached the elective replacement indicator.The atrial lead was capped on (b)(6) 2019 due to unresolvable noise and the left ventricular lead was capped on (b)(6) 2019 because it was non-responsive.
 
Event Description
New information revealed that the icd experienced normal battery depletion.The lv lead was performing optimally, but the patient's ejection fraction remained 20%.The physician elected to place the new lv lead in a different vessel and use a his lead to improve ef.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8985199
MDR Text Key157353488
Report Number2938836-2019-13439
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2015
Device Model Number1458Q/86
Device Lot Number3838088
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/17/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/17/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-