Model Number 5392
Device Problems Failure to Power Up (1476); Contamination /Decontamination Problem (2895)
Patient Problem No Patient Involvement (2645)
Event Date 07/24/2019
Event Type  malfunction  
Manufacturer Narrative
Product analysis : analysis confirmed customer comment that the external pulse generator would not turn on.The device had internal contamination all over the place.The hanger assembly is cracked.All found defective parts were replaced and all other identified issues were resolved.Passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) would not turn on.The epg was returned for service.There was no patient involvement.It was further reported upon analysis that the epg had contamination all over the inside of the unit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8996840
MDR Text Key158813923
Report Number3004593495-2019-00931
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00643169725126
UDI-Public00643169725126
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2019
Date Manufacturer Received08/13/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/25/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-