Model Number 3664
Device Problem Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem Inadequate Pain Relief (2388)
Event Date 04/01/2019
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
 
Event Description
It was reported the ipg was not placed in surgery mode prior to the patient undergoing an unrelated surgical procedure in (b)(6) (exact date unknown).Postoperatively, the ipg would no longer communicate with external devices, and the patient controller displayed the error message,.
 
Manufacturer Narrative
Updated from malfunction to serious injury.The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
 
Event Description
Additional information received which indicated that patient underwent surgical intervention wherein the ipg was replaced on (b)(6) 2020 and effective therapy was restored post operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM DRG IPG
Type of Device
DRG IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key9029189
MDR Text Key158913824
Report Number1627487-2019-10226
Device Sequence Number1
Product Code PMP
UDI-Device Identifier05415067020215
UDI-Public05415067020215
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 06/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/19/2019
Device Model Number3664
Device Catalogue Number3664
Device Lot Number6133812
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/22/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/06/02/2017/001-C
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight79
-
-