Catalog Number MMC9609L |
Device Problem
Complete Blockage (1094)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/15/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Lot number - the customer reported a potentially associated lot number, 18k13t426.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a blood set had clots during priming.The reporter stated that when priming the set for a blood transfusion, large clots were noticed in the chamber of the set.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information :lot number provided: 18k13t426 : the device was received contaminated with blood.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional flow test was performed and liquid flows through appropriately.The reported problem was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|