Additional information : correction to the previously submitted date.Baxter aware date is august 21, 2019 previously submitted as august 20, 2019.The device was received for evaluation.Visual inspection was performed and observed that the inner bag of the set sticks to the outer packaging.The reported condition was verified.The cause of the condition was found to be related to a human production issue during manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|