• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - TUNISIA TUR IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - TUNISIA TUR IRRIGATION SET; SYSTEM, IRRIGATION, UROLOGICAL Back to Search Results
Catalog Number VMC4005
Device Problem Packaging Problem (3007)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
This event occurred during the unspecified date of (b)(6) 2019.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the uromatic y type tur set "sticks to the outer packaging so the product cannot be taken out of the package in a sterile manner".This was identified prior to patient use; therefore, there was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information : correction to the previously submitted date.Baxter aware date is august 21, 2019 previously submitted as august 20, 2019.The device was received for evaluation.Visual inspection was performed and observed that the inner bag of the set sticks to the outer packaging.The reported condition was verified.The cause of the condition was found to be related to a human production issue during manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUR IRRIGATION SET
Type of Device
SYSTEM, IRRIGATION, UROLOGICAL
Manufacturer (Section D)
BAXTER HEALTHCARE - TUNISIA
tunis
Manufacturer (Section G)
BAXTER HEALTHCARE - TUNISIA
route de chebbaou
2021oued e
tunis
TS  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9039395
MDR Text Key158883385
Report Number1416980-2019-05027
Device Sequence Number1
Product Code LJH
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 10/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberVMC4005
Device Lot Number18K10T431N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/24/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-