GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH Back to Search Results
Model Number FGS-0450
Device Problems Loss of Power (1475); Unintended Application Program Shut Down (4032)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/14/2019
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: this report is based on information provided by medtronic investigation personnel and the sample that arrived.One bravo capsule and one bravo delivery device were received for evaluation.A review of the product expiration date discovered this product was used before the expiration date.The returned sample met specification as received by medtronic.The visual inspection found no notable conditions.The customer reported bravo fail to attach to patient's esophagus.The investigation of the returned equipment did not identify anything that would have caused or contributed to the reported event.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the recorder turned off after only 25 minutes.Technical support connected to the system and confirmed a short study.There was no patient and user harm.A repeat procedure on a different day was necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO
Type of Device
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer (Section G)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer Contact
amy beeman
161 cheshire lane, suite 100
plymouth, MN 55441
7632104064
MDR Report Key9072933
MDR Text Key158819446
Report Number9710107-2019-00419
Device Sequence Number1
Product Code FFT
UDI-Device Identifier07290101363064
UDI-Public07290101363064
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFGS-0450
Device Catalogue NumberFGS-0450
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2019
Date Manufacturer Received08/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/13/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age71 YR
Patient Weight86
-
-