• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VENTANA MEDICAL SYSTEMS BENCHMARK ULTRA STAINER MODULE; SLIDE STAINER, AUTOMATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENTANA MEDICAL SYSTEMS BENCHMARK ULTRA STAINER MODULE; SLIDE STAINER, AUTOMATED Back to Search Results
Catalog Number 05342716001
Device Problem Fluid/Blood Leak (1250)
Patient Problem Fall (1848)
Event Date 08/22/2019
Event Type  malfunction  
Manufacturer Narrative
The local field service engineer cleaned the clogged waste filter; this service action remedied the customer's observation.A new waste tub and filter kit have been released to increase the reliability of the waste system.This is achieved by increasing the holding capacity of the waste tub and removing the inline waste filter.A new customer cleanable waste strainer has been created and added to the waste tub.(b)(4).
 
Event Description
A customer in (b)(6) reported that their benchmark ultra stainer module leaked, and that the leakage reached the floor.No harm or injury occurred as a result of the leakage and fluid on the lab's floor.The following failure mode is same event reported in mdr-2028492-2016-00004-00, which caused a slip and fall injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENCHMARK ULTRA STAINER MODULE
Type of Device
SLIDE STAINER, AUTOMATED
Manufacturer (Section D)
VENTANA MEDICAL SYSTEMS
1910 e. innovation park dr.
tucson AZ 85755
Manufacturer Contact
stacie-ann creighton
1910 e innovation park dr
na
tuscon, AZ 85755
9082537112
MDR Report Key9079129
MDR Text Key202057517
Report Number2028492-2019-00018
Device Sequence Number1
Product Code KPA
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
510K EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2019
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05342716001
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/22/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-