TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
|
Back to Search Results |
|
Model Number 164267 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/26/2019 |
Event Type
malfunction
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the unit displayed an "under speed" error message.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.The service repair technician (srt) duplicated the reported complaint.The cable assembly was replaced and the unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Manufacturer Narrative
|
Updated blocks: d10, h3, and h6.During laboratory analysis, the product surveillance technician (pst) observed the centrifugal drive motor to display an "underspeed error message".It was examined that there was a breach in the cable jacket around the strain relief area causing a short.Voltage fluctuations occurred when the cable was manipulated determining that to be the cause of the underspeed error message.
|
|