• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS CENTRIFUGAL SYSTEM; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM Back to Search Results
Model Number 164267
Device Problem Output Problem (3005)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/26/2019
Event Type  malfunction  
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the unit displayed an "under speed" error message.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.The service repair technician (srt) duplicated the reported complaint.The cable assembly was replaced and the unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Manufacturer Narrative
Updated blocks: d10, h3, and h6.During laboratory analysis, the product surveillance technician (pst) observed the centrifugal drive motor to display an "underspeed error message".It was examined that there was a breach in the cable jacket around the strain relief area causing a short.Voltage fluctuations occurred when the cable was manipulated determining that to be the cause of the underspeed error message.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS CENTRIFUGAL SYSTEM
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key9080957
MDR Text Key159611444
Report Number1828100-2019-00505
Device Sequence Number1
Product Code DWA
UDI-Device Identifier00886799000458
UDI-Public(01)00886799000458(11)170727
Combination Product (y/n)N
PMA/PMN Number
K950739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 12/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/18/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number164267
Device Catalogue Number164267
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2019
Was the Report Sent to FDA? No
Date Manufacturer Received11/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-