The evaluation was completed.The instrument was examined under 45x magnification.There was no metallic debris seen on the part.The scissors and most particularly the space behind the scissor in between the arms was coated with dried surgical debris.There were no areas identified on the part that looked like they may have been damaged or had metal particulate present.The disposable tip was examined under magnification typically used during the manufacture of the device.There were no worn areas, slivers or burrs seen on the instrument.There was some evidence of corrosion seen on the instrument on the back end of the d-tip.The report was not confirmed as there was no metal debris or indications that any metal debris was generated from the disposable scissor seen in the evaluation of the unit returned.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action required.
|